Medtronic Paradigm 522 Guide de l'utilisateur Page 97

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 262
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 96
2 Select Carb Ratios, then press ACT.
If you use gr
ams as your carb units: Carb ratio is the number of carb grams that are covered by
one (1.0) unit of insulin.
If you use exchanges as your carb units: Carb ratio is the number of insulin units that are needed
to cover one (1.0) carb exchange.
3 The SET CARB RATIO 1 (if you are using grams) or SET EXCH RATIO 1 (if using exchanges) screen
appears. The default ratio flashes on the screen.
4 Set your first ratio, then press ACT. Carb ratio values are normally between 5-50 grams/u or 0.3-3.0
u/exch. If your ratio value is outside the range, a warning message will appear on the screen. This
message warns that the entered carb ratio is valid but outside the usual range. Press ESC to correct or
ACT to continue.
The start time for your first ratio is midnight and cannot be changed.
5 The SET START TIME 2 screen appears. The dashes under the screen name flash. The first Carb Ratio or
Exchange Ratio is now set.
If you do not need a second ratio, press ESC and skip to the next section. If you need to set up another
ratio, follow steps 6 through 9.
6 In the SET START TIME 2 screen, enter the time of the day you want this ratio to become active.
7 Press ACT. The SET CARB RATIO 2 screen (if using grams) or SET EXCH RATIO 2 screen (if using
exchanges) appears.
8 The default ratio flashes. Select your ratio.
9 Press ACT. The SET START TIME 3 screen appears. The second carb or exchange ratio is now set.
10 If you do not need to set up any more ratios, press ESC. If you need to set up more ratios, repeat steps
6 through 9 above for each ratio.
You are now ready to set up the BG units.
Setting the BG units
Y
ou can select mmol/L or mg/dL as your Blood Glucose Unit (measurement type). If you make changes to
your BG settings, you must reprogram your insulin sensitivity and BG targets.
1 Make sure the EDIT SETTINGS screen is open.
Main > Bolus > Bolus Wizard Setup > Edit Settings
2 Select BG Units, then press ACT.
3 The BLOOD GLUCOSE UNITS screen displays. Select mmol/L or mg/dL, then press ACT.
4 The EDIT SETTINGS screen shows the BG units you selected. You are now ready to set your insulin
sensitivity
.
Using the Bolus Wizard feature78