2 With the tester arms pinched, gently pull the transmitter away from the tester. To save transmitter
battery life, do not leave the tester connected after testing.
Troubleshooting sensor features
Reconnect old sensor
You should only use this feature if you have disconnected the sensor from the transmitter and have to
reconnect them (for example, when flying on an aircraft).
To Reconnect old sensor:
1 Go to the RECONNECT OLD SENSOR screen.
Main > Sensor > Sensor Start > Reconnect Old Sensor
2 Press ACT in the
RECONNECT OLD SENSOR screen.
3 The SENSOR READY 2 HRS screen appears. Press any button to continue. An alert will sound in two
hours when a meter BG is required for calibration.
Find lost sensor
If you receive a Lost Sensor alert:
1 Relocate the pump closer to the sensor, and go to the SENSOR START MENU.
Main > Sensor > Sensor Start
2 Select Find Lost Sensor, then press ACT.
3 The SENSOR READY 15 MIN screen appears. Press any button to continue. An alert will sound in 15
minutes when a meter BG is required for calibration.
Troubleshooting and alarms 169