Areas to avoid:
Frequently used injection or pump/sensor sites
2-inch (5.0 cm) area around navel
Site where clothing will rub or constrict
Scarred or hardened tissue
Areas subjected to a lot of movement
Clean site with alcohol, making sure site is dry before inserting the sensor. Do not use skin-preparation
solutions prior to insertion. However, I.V. Prep may be used after insertion and before applying a sterile
dressing. Lift back of tape slightly to apply I.V. Prep.
Always refer to the instructions that shipped with your glucose sensor.
1 Wash your hands.
2 Clean the sensor site with alcohol. Let it air dry.
3 Remove the sensor from the package by holding the base or tape. Do not hold the sensor by the
introducer needle handle.
4 Place the sensor in the Sen-serter until it fits snugly.
5 Holding the white tape as shown, remove the clear tape using a counterclockwise motion.
6 Place your fingers on the back of the white tape and push the carrier down until it clicks into place.
7 Turn the white button to lock the Sen-serter. Remove the needle guard from the introducer needle.
8 Rest the Sen-serter legs flat on your skin so the Sen-serter is at a 45-degree angle or greater
to the
insertion site. Bleeding can occur if you insert the sensor at an angle less than 45 degrees. Place two
fingers of your other hand on the Sen-serter legs to maintain the correct angle.
9 Turn the white button to unlock the Sen-serter. Press the white button to insert the sensor.
10 Make sure the sensor is inserted. If the sensor is not fully inserted, manually push it into place. The
sensor should lay flat on your skin.
Sensor features120