Starting the sensor
T
o start the sensor working, you must complete the following steps in order:
Insert a battery into the transmitter's charger.
Charge the transmitter battery.
Set up the sensor features.
Insert the sensor and wait five minutes.
Connect the transmitter to the sensor.
Start the sensor and wait two hours.
Enter your first meter BG.
Inserting the sensor
Before inserting the sensor, you must fully charge and set up the transmitter. Also, if the sensor has been
refrigerated, remove the sensor package from refrigeration. To avoid condensation, make sure that you
allow the sensor package to reach room temperature before opening.
The sensor is inserted through the skin with an insertion device called the Sen-serter
®
and placed in the
fatty layer under the skin. The sensor produces a signal that reflects the amount of glucose in the
interstitial fluid at the insertion site. This signal is sent to the transmitter, which is then sent to the pump.
The pump translates the signal and displays a sensor glucose reading on your pump screen.
WARNING: The sensor may create special needs regarding your medical conditions or medications.
Please discuss these conditions and medications with your doctor before using the sensor
.
Bleeding, swelling, irritation and/or infections at the insertion site are possible risks associated with
inserting the sensor and may result from improper insertion and maintenance of insertion site.
Choose a site with an adequate fatty layer for sensor insertion. Be sure to
rotate the sensor sites so that they do not become overused. Shown here are
the best body areas (shaded) for sensor insertion. The clinical trials of sensor
accuracy have been based on sensors inserted in the abdominal area. Sensor
performance may differ when other insertion sites are used.
CAUTION: Never insert the sensor within 2 inches (5.0 cm) from the
infusion set insertion site or within 3 inches (7.5 cm) from the manual
injection site.
Sensor features 119
Chapter 8