Active insulin time
The active insulin time setting lets the pump know which active insulin time to use in calculating the
amount of active insulin to subtract before estimating a bolus. Y
our healthcare professional should
determine the active insulin time that is best for you.
To set the Active insulin time, do the following steps:
1 Make sure the EDIT SETTINGS screen is open.
Main > Bolus > Bolus Wizard Setup > Edit Settings
2 Select Active Ins T
ime, then press ACT.
3 The ACTIVE INS TIME screen appears. The default time of six hours flashes on the screen.
4 Set the number of hours for the active insulin time, then press ACT.
5 The EDIT SETTINGS screen shows the new Active insulin time setting. You have now completed the
Bolus Wizard setup. Press ESC or wait to see the message: Bolus Wizard setup is complete.
Review your Bolus Wizard feature settings
Check your Bolus W
izard feature settings in the REVIEW SETTINGS screen. If necessary, compare this
information with your information in the Bolus Wizard feature settings table.
1 Go to the REVIEW SETTINGS screen.
Main > Bolus > Bolus Wizard Setup > Rev
iew Settings
2 Scroll through the text to view your Bolus Wizard settings.
3 Exit the menus when you are done.
Using the Bolus Wizard feature82