Auto-off
The factory setting for this feature is Off. This is a safety feature that stops insulin delivery after a
defined time period (from 1 to 24 hours). If the pump detects that no buttons have been pressed for the
selected amount of time in Auto-off, insulin delivery will stop and an alarm will sound. Y
ou may choose to
program this feature into your pump based on the number of hours that you usually sleep at night. Discuss
what uses and settings are best for you with your healthcare professional.
1 Go to the AUTO OFF DURATION screen.
Main > Utilities > Alarm > Auto Off
2 Set the number of hours you want to set and press ACT.
If you do not want to use the Auto Off feature, make sure the hour is set to zero (0).
3 The screen will return to the ALARM MENU. The Auto Off feature is now set. Exit the menus.
LOW RESV WARNING (Low reservoir warning)
Allows you to program the pump to sound an alert before your reservoir is empty. You can select one of
these warning types:
a specified number of units that remain in the reservoir
a specified maximum amount of time that remains before the reservoir will be empty
The factory setting for this feature is 20 insulin units.
If you use Time as the low reservoir warning type and you deliver large boluses, the actual time remaining
could be less than the warning time. The warning time is based on the basal insulin delivery rate. The
Time low reservoir warning type is intended to let you know if you will have enough insulin while you are
sleeping.
1 Go to the RESV WARNING TYPE screen.
Main > Utilities > Alarm > Low Resv W
arning
2 Select Insulin Units or Time and press ACT.
WARNING: When the pump detects a low reservoir condition during a bolus or prime delivery,
the LOW RESERVOIR alert displays after the delivery is finished. Make sure to check the
volume of your reservoir to ensure enough insulin is available.
Utilities136