Pump motor
The pump motor has a unique, patented design with an integrated safety check system. The system
delivers in precise increments.
Dual W
ave bolus
Delivers a Normal bolus followed by a Square Wave bolus (limited by maximum bolus).
Easy bolus
Features programming using audible tones (or vibrate pulses) in user-defined increments.
Beep mode range: 0 to maximum bolus
Vibrate mode range: 0 to 20 steps or maximum bolus, whichever comes first
Default step increment: 0.1 unit
Step size < maximum bolus.
Adjustable step size: 0.1 to 2.0 units per step.
Accessible from the remote control or pump buttons.
Infusion pressure
Maximum infusion pressure and occlusion pressure: 13.7 PSI (94.46 kPa).
Insulin sensitivity
Maximum settings: 8
Factory default: 50 mg/dL (2.8 mmol/L)
Range: 10 - 400 mg/dL (0.5 - 22.2 mmol/L)
Warning limits: less than 20 or greater than 100 mg/dL (less than 1.1 or greater than 5.6 mmol/L)
Pump specifications 181