Medtronic Paradigm 522 Guide de l'utilisateur Page 215

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 262
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 214
Sensor features default settings
Feature Options Default Setting Limits, Range of Values
Sensor
On
Off
Off None
High Glucose
Alert
On
Off
Off Low–22.2 mmol/L (Low–400
mg/dL)
a
Low Glucose
Alert
On
Off
Off 2.2 mmol/L–High
(40mg/dL–High)
b
Alarm Snooze (Always on - cannot be turned
off)
On 30 minutes Range: 0:05 - 1:00 (5
minutes - 1 hour)
Cal Reminder
On
Off
On 1 hour Range: 0:05 - 4:00 (5
minutes - 4 hours)
Blood Glucose
Units (BG Units)
mg/dL
mmol/L
mg/dL None
Missed Data (Always on - cannot be turned
off)
30 minutes Range: 0:05 - 0:40 (5
minutes - 40 minutes)
High Snooze 1:00 (1 hour) Range: 0:05 - 3:00 (5
minutes - 3 hours)
Low Snooze 0:20 (20 minutes) Range: 0:05 - 1:00 (5
minutes - 1 hour)
a. Your High Glucose limit must be at least 0.6 mmol/L (10 mg/dL) above your Low Glucose limit.
b. Your Low Glucose limit must be at least 0.6 mmol/L (10 mg/dL) below your High Glucose limit. For example, if your High
Glucose limit is 11.1 mmol/L (200 mg/dL), then your Low Glucose limit cannot be higher than 10.5 mmol/L (190 mg/dL).
Pump specifications196
Vue de la page 214
1 2 ... 210 211 212 213 214 215 216 217 218 219 220 ... 261 262

Commentaires sur ces manuels

Pas de commentaire