Medtronic Paradigm 522 Guide de l'utilisateur Page 210

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 262
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 209
#3: No active insulin from previous bolus delivery. User inputs 60 grams of carbs and a BG of 70 mg/dL (3.9
mmol/L).
(food estimate)
60 g
30 g/u
= 2 units
= -0.5 units
+
(correction estimate)
= 2 + (-0.5)
= 1.5 units
bolus estimate
3.9 mmol/L - 5.0 mmol/L
2.2 mmol/L/u
= -0.5 units
70 mg/dL - 90 mg/dL
40 mg/dL/u
#4: No active insulin from previous bolus delivery. User inputs 60 grams of carbs and a BG of 100 mg/dL
(5.6 mmol/L).
(food estimate)
60 g
30 g/u
= 2 units
+
(correction estimate)
= 2 + 0
= 2 units
bolus estimate
Correction is 0 because the
current blood
glucose reading
is between the BG High and
Low Targets
Pump specifications 191
Chapter 15
Vue de la page 209
1 2 ... 205 206 207 208 209 210 211 212 213 214 215 ... 261 262