Pump alert conditions
Listed below are alerts that indicate a condition that is outside normal pump activity
.
LOW RESERVOIR
You can program the pump to sound an alert when either a specified number of units remains or a
specified amount of time remains before the reservoir will be empty.
LOW BATTERY
If you get this alert, do not go to sleep without replacing the battery. The backlight, the remote, and
meter functions are disabled during a LOW BATTERY condition. If the alert type is set to Vibrate, the pump
will change to the audio alert Beep-medium. Clear (ESC, ACT) this alert before you replace your battery.
Sensor alert conditions
Listed below are the alerts that you may encounter while using the sensor feature of your pump, and how
to resolve the alert condition.
WEAK SIGNAL
Reason: Occurs when the pump does not receive data from the transmitter for a pre-defined period of
time (as set in Missed Data).
Action: Move the pump closer to the transmitter or move the transmitter and the pump to the same side
of your body.