Medtronic Paradigm 522 Guide de l'utilisateur Page 73

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 262
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 72
If you use patterns, the BASAL REVIEW screen appears.
The current basal pattern will be highlighted. Select the pattern you want to view
. Press ACT.
The delivery details for that pattern will appear.
4 Exit the menus when you are done.
Setting the Max basal rate
Maximum basal (Max basal) rate is a safety limit for the amount of basal insulin that is able to be delivered
per hour
. Your pump is sent from the factory with the Max basal set to 2.0 units per hour. It is important to
determine your Max basal rates with your healthcare professional. This safety feature will not allow you to
program any basal rates, including patterns and temporary basal rates, that are greater than the Max basal
rate.
If you are setting your Max basal rate after your basal rates have been set, you cannot set a Max basal that
is less than any of the programmed basal rates.
To set your Max basal rate, do these steps:
1 Go to the MAX BASAL RATE screen. The Max basal rate will be flashing.
Main > Basal > Max Basal Rate
2 Change the rate and press ACT.
3 Your Max basal rate is now set. Exit the menus.
Example 1: Max basal
Helen has a very low insulin requirement. Her highest basal rate is only 0.4 units per hour. As a safety
measure, Helen's healthcare professional set her pump with a Maximum basal rate of 1.0 units per hour
.
Example 2: Max basal
Rusty needs large amounts of insulin to control his blood glucose levels. His new pump was delivered
from the factory with a Maximum basal rate of 2.0 units per hour
, but he needs 2.8 units per hour in the
early morning. Rusty will reprogram his Maximum basal to 3.0 units per hour to accommodate his needs.
Basic programming54
Vue de la page 72
1 2 ... 68 69 70 71 72 73 74 75 76 77 78 ... 261 262