Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The Paradigm 522/522K/722/722K pump system is intended for use in the electromagnetic
env
ironment specified below. The customer or user of the Paradigm 522/522K/722/722K pump
system should make sure that it is used in such an environment.
Immunity
Test IEC 60601
Level
Compliance
Level Electromagnetic Environment Guidance
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the Paradigm 522/722/522K/722K pump is used exceeds the applicable RF compliance
level above, the Paradigm 522/722/522K/722K pump should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as re-orienting or
relocating the Paradigm 522/722/522K/722K pump.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V
1
] V/m.
Pump specifications202