Medtronic Paradigm 522 Guide de l'utilisateur Page 143

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 262
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 142
Entering meter BG
Sensor calibration will be successful only if the BG entry is in the range of 40-400 mg/dL (or 2.2 to 22
mmol/L). Y
ou should calibrate three to four times spread throughout the day for optimal results.
1 Use one of these ways for all the calibrations throughout the sensor's life, including the first
calibration after the initialization.
a. T
o enter a meter BG value manually, follow this path:
Main > Sensor > Enter Meter BG
b. If at the time of sensor calibration you also need to perform a bolus using the Bolus Wizard, you
can use the meter BG for both calibrating the system and bolusing. To enter a meter BG value
manually in the Bolus Wizard, press the
button on your pump, or use the MAIN MENU to go to
the
ENTER BG screen:
Main > Bolus > Use Bolus Wizard
c. To enter a BG value automatically through the linked meter, test your BG with a fingerstick. The
value will be automatically sent from the meter to your pump. Make sure that your pump and the
meter are properly programmed for communication. See the Meter option section in the Using the
Bolus Wizard feature chapter for more information.
2 If you are manually entering your meter BG, use the up and down arrow buttons on your pump to do
so. P
ress ACT. For the Bolus Wizard option above, select Yes when the following screen displays: BG
TO UPDATE SENSOR
Sensor features124
Vue de la page 142
1 2 ... 138 139 140 141 142 143 144 145 146 147 148 ... 261 262

Commentaires sur ces manuels

Pas de commentaire