Delivery accuracy
•
Delivery accuracy: +/-5 percent.
• Displacement (piston) accuracy: +/-2 percent.
•
The delivery accuracy test was done using the Sof-Set
®
Ultimate infusion set (MMT–315).
-50
-40
-30
-20
-10
0
10
20
30
40
50
0 15 30 45 60 75 90 105 120 135 150 165 180 195 210
Maximum Error (percent)
Minimum Error (percent)
Ovreall Flow Error (percent)
Time Interval (Minutes)
Percent Error
Delivery Volume Accuracy Curve (-0.37%) Overall Error
(MMT-522 Model) Qualification Testing (0.01 mL/hr 09/22/04 S/N: 010112-C061); 23°C
Pump specifications180