NO DELIVERY
Insulin delivery has stopped. This alarm occurs if your pump detects a blockage.
NO RESERVOIR
The reservoir is not inserted correctly or no reservoir has been inserted.
OFF NO POWER
The battery is dead. Replace battery immediately. Follow the directions on the screen. Check to make
sure that the time is correct on the screen. Reset the time if necessary.
RESET
The Reset alarm triggers when pump settings are cleared because of one of these reasons:
• Pump settings were cleared (Clear Settings function) and settings have not been reprogrammed.
• A download attempt from the PC is incomplete. (The download function is applicable to the optional
software feature. Refer to the software user guide for more information.)
WEAK BATTERY
The pump tests the voltage of each battery installed. If the battery voltage is less than full strength, this
alarm may occur. The pump will operate normally, but the battery life will be shorter than expected.
Always make sure to install a new battery in the pump.
Testing the transmitter
The tester acts like a sensor. If you get sensor-related alerts, use the tester to make sure that the
transmitter is working.
Connecting the tester
1 Hold the transmitter and the tester as shown. Line up the flat side of the tester with the flat side of
the transmitter
.
Troubleshooting and alarms 167
Chapter 13
Commentaires sur ces manuels