Rewinding your pump
Before you continue, make sure the infusion set is NOT connected to your body and the reservoir is NOT in
the pump.
WARNING: Make sure the infusion set is disconnected from your body before you rewind your pump or
fill the infusion set tubing. Never insert the reservoir into the pump while the tubing is connected to
your body. Doing so could result in an accidental infusion of insulin.
1
ಘ
If you removed your reservoir and are replacing it, go to the REWIND screen.
Main > Prime > Rewind
2
ಘ
In the REWIND screen, press ACT to start the rewind process. The REWINDING screen will appear while
the pump rewinds.
3
ಘ
After the pump rewinds the PREPARING TO PRIME screen will appear.
If you are practicing:
a. Do NOT insert the reservoir in your pump. Make sure the red shipping cap is installed in the reservoir
compartment as shipped.
b. Press ACT, then continue with the manual prime instructions described in the Manual prime section
of this chapter.
If you are not practicing, continue to the next section to insert the reservoir in your pump.
Inserting the reservoir in your pump
If your reservoir is already inserted in your pump, continue to the next section.
You must do these steps in the order described. If you are practicing, do NOT insert the reservoir in your
pump.
CAUTION: You must rewind your pump before installing a new reservoir. As part of the pump's function,
it calculates the reservoir volume. To ensure correct volume calculations, your pump has been designed
to require a rewind before you insert your reservoir.
Starting on insulin
63