4
ಘ
When charging is complete, the green light on the charger will stay on, without flashing, for 15 to
20 seconds and then turn off.
5
ಘ
After the green charger light turns off, disconnect the transmitter from the
charger. The green light on the transmitter will flash for approximately five
seconds and then turn off.
ಘ
6
ಘ
If the green light on the transmitter does not flash, reconnect it to the charger for at least one minute.
7
ಘ
After removing the transmitter from the charger, wait at least one minute before connecting it to a
sensor or tester.
Starting the sensor
To start the sensor working, you must complete the following steps in order:
• Insert a battery into the transmitter's charger.
• Charge the transmitter battery.
• Set up the sensor features.
• Insert the sensor and wait five minutes.
• Connect the transmitter to the sensor.
• Start the sensor and wait two hours.
• Enter your first meter BG.
Inserting the sensor
Before inserting the sensor, you must fully charge and set up the transmitter. Also, if the sensor has been
refrigerated, remove the sensor package from refrigeration. To avoid condensation, make sure that you allow
the sensor package to reach room temperature before opening.
The sensor is inserted through the skin with an insertion device called the Sen-serter
®
and placed in the fatty
layer under the skin. The sensor produces a signal that reflects the amount of glucose in the interstitial fluid
at the insertion site. This signal is sent to the transmitter, which is then sent to the pump. The pump
translates the signal and displays a sensor glucose reading on your pump screen.
Sensor features
121
ಘChapter 8