Sensor features
The optional sensor and transmitter can provide continuous glucose measurements to help you control your
glucose levels better. The sensor measures the glucose levels in the fluid under your skin. The transmitter
receives this measurement from the sensor and sends it wirelessly to the pump. To take advantage of your
pump sensor features, call 800-646-4633, +1-818-362-5958 (outside U.S.), or visit our web site at
www.medtronicdiabetes.com to purchase your sensor and transmitter.
The sensor glucose and meter blood glucose measurements are not identical. To see the accuracy of the
sensor glucose measurements, see the Sensor accuracy appendix.
Entering your sensor settings
Enter your pump sensor settings in the order they appear below, as some of the settings are dependent upon
other settings being made first. When you complete each setting, your pump will automatically display the
screen for the next setting in the sequence. When a screen item flashes, you may change the value of the
flashing item by pressing
or .
Turning on the sensor
The sensor must be turned on, started and initialized to report glucose measurements.
1
Go to the EDIT SETTINGS screen.
Main > Sensor > Sensor Setup > Edit Settings
The EDIT SETTINGS screen appears with Sensor: Off selected.
2
Press ACT. The SENSOR ON/OFF screen appears.
3
Select On, then press ACT. The EDIT SETTINGS screen appears showing the sensor is now turned on.
You are now ready to set up the sensor features.