Medtronic Paradigm Link Guide de l'utilisateur Page 238

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 262
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 237
Effects of calibration frequency
The average bias when Guardian RT was calibrated ~ 3.5 times a day was -20.5 ± 41 mg/dl (LL: -22.40 mg/dl
and UL: -18.63 mg/dl). In those calibrated ~ 5 times a day, the bias was -10.2 mg/dl ± 36 mg/dl (LL: -11.74
mg/dl and UL: -8.66 mg/dl). When comparing Guardian RT units that were calibrated less often to those
calibrated more often, the following alarm performance was observed:
Specificity increased 2-4 percent in the hypoglycemic range and decreased 0-2 percent in the
hyperglycemic range
Sensitivity increased between 5-9 percent across the hyperglycemic range, and decreased 7-16 percent
when the alarm was set to 80 mg/dl or below, and decreased 3-7 percent when set between 85 and 100
mg/dl
Stratified error grid analysis also shows better performance in the hypoglycemic range when fewer
calibrations are performed, i.e., 62 percent of data points are in Zone A when fewer calibrations were
performed, whereas 58 percent were in Zone A when more calibrations were performed.
Sensor accuracy
219
Appendix A
Vue de la page 237
1 2 ... 233 234 235 236 237 238 239 240 241 242 243 ... 261 262

Commentaires sur ces manuels

Pas de commentaire