Selecting the BG units
You must select the blood glucose units (BG Units) for the system (either mmol/L or mg/dL). All BG
measurements will show in the BG unit type you select.
If you have turned on your Bolus Wizard feature, BG units will not be available for sensor setup. Please go
to the Setting the BG units section of the Using the Bolus Wizard feature chapter for details.
1
Make sure the EDIT SETTINGS screen is open.
Main > Sensor > Sensor Setup > Edit Settings
2
Select BG Units, then press ACT. The BLOOD GLUCOSE UNITS screen shows mg/dL and mmol/L.
3
Select your BG units.
4
Press ACT. The EDIT SETTINGS screen shows the BG Units type you selected.
You are now ready to set up your High Glucose limit.
High Glucose Alert
Your High Glucose alert must be turned on if you want your pump to alert when the sensor glucose
measurement reaches or goes above your set High Glucose limit.
To turn on your High Glucose alert, and to set the High Glucose limit:
1
Make sure the EDIT SETTINGS screen is open.
Main > Sensor > Sensor Setup > Edit Settings
2
Select High Glucose, then press ACT.
3
Select On, then press ACT. The SET HIGH GLUCOSE screen flashes the default High Glucose limit of 200
mg/dL (11.2 mmol/L).
4
Select your High Glucose limit between 110–400 mg/dL (2.8–22.2 mmol/L). Your High Glucose limit must
be at least 10 mg/dL (0.6 mmol/L) above your Low Glucose limit.
5
Press ACT. The EDIT SETTINGS screen shows the High Glucose value you selected. You are now ready to
set up your High Snooze.
Setting the High Snooze
After you receive and clear a High Glucose alert, it will repeat until the high glucose condition is resolved.
The High Snooze feature allows you to set how frequently you want the High Glucose alert to repeat after
you clear it the first time. For example, if your blood glucose levels usually take one to two hours to fall
after a correction bolus, you may want to set the High Snooze to one or two hours to avoid unnecessary alerts.