Medtronic Paradigm Link Guide de l'utilisateur Page 73

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 262
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 72
3
If you do not use patterns, the
STANDARD screen appears.
The delivery details for your standard
basal will appear.
If you use patterns, the BASAL REVIEW
screen appears.
The current basal pattern will be
highlighted. Select the pattern you want
to view. Press ACT.
The delivery details for that pattern will
appear.
basal rate
start times
basal
delivery rates
(24-hour total)
basal insulin
basal rate
start times
basal
delivery rates
(24-hour total)
basal insulin
4
Exit the menus when you are done.
Setting the Max basal rate
Maximum basal (Max basal) rate is a safety limit for the amount of basal insulin that is able to be delivered
per hour. Your pump is sent from the factory with the Max basal set to 2.0 units per hour. It is important to
determine your Max basal rates with your healthcare professional. This safety feature will not allow you to
program any basal rates, including patterns and temporary basal rates, that are greater than the Max basal
rate.
Basic programming