Recommended separation distances between portable and mobile RF communications
equipment and the Paradigm 522/722/522K/722K pump system
This section provides information on the recommended separation distance between portable and
mobile RF communications equipment and the Paradigm 522/722/522K/722K pump. The Paradigm
522/722/522K/722K pump is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. Paradigm 522/722/522K/722K pump users can help prevent electro-
magnetic interference by maintaining a minimum distance between portable and mobile RF commu-
nications equipment (transmitters) and the Paradigm 522/722/522K/722K pump as recommended
below, according to the maximum output power of the communications equipment.
Separation distance according to the frequency of transmitter (m)
Rated maximum out-
put power of trans-
mitter (W)
80MHz to 800MHz
d = 0.35
P
800MHz to 6.0GHz
d = 0.70
P
0.01 0.035 0.07
0.1 0.11 0.11
1 0.35 0.7
10 1.1 2.2
100 3.5 7
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where
p is the maximum output power rating of the transmitter in watts (W) according to the transmitter man-
ufacturer.
NOTE: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.