Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The Paradigm 523/723/523K/723K pump system (consisting of the MMT-523/723/523K/723K pump
and MMT-7703 transmitter) is intended for use in the electromagnetic environment specified below.
The customer or the user of the Paradigm 523/723/523K/723K pump system should make sure that
it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance
Level
Electromagnetic Environment
- Guidance
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The Paradigm 523/723/523K/723K pump is intended for use in the electromagnetic environment
specified below. The customer or user of the Paradigm 523/723/523K/723K pump system should
make sure that it is used in such an environment.
Immunity
Test IEC 60601
Level
Compliance
Level Electromagnetic Environment Guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the
Paradigm 523/723/523K/723K pump, including
cables, than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
Conducted RF
IEC 61000-4-6
10 Vrms
150 kHz to
80 MHz
Not applicable Not applicable
Pump specifications200