Bolus Wizard feature default settings
Item Default
Setting
Limits Increments Warning Limits
carb units: grams
ins to carb (or
exch) ratio:
15 grams/u
or
1 unit/exch
1–200 g/u or 0.075–
15.0 u/exch
0.1 g/u for 0–9.9 g/u;
1 g/u for ratios of 10.0 g/u
or more
(or 0.001 u/exch for 0.075–
0.099 u/exch; 0.01 u/exch
for 0.10–9.99 u/exch;
0.1 u/exch for 10.0–
15.0 u/exch)
5–50 g/u or 0.3–
3.0 u/exch
(insulin)
sensitivity:
50 mg/dL/u
or
2.8 mmol/L/
u
10–400 mg/dL/u or
0.5–22.2 mmol/L/u
1 mg/dL/u or
0.1 mmol/L/u
20–100 mg/dL/u
or 1.1–
5.6 mmol/L/u
BG Target: 100–
100 mg/dL or
5.6–
5.6 mmol/L
60–250 mg/dL or
3.3–13.9 mmol/L
1 mg/dL or 0.1 mmol/L 90–140 mg/dL or
5.0–7.8 mmol/L
Active Ins
Time:
6 hours 2–8 hours 1 hour
Sensor features default settings
Feature Options Default Setting Limits, Range of Values
Sensor
•On
•Off
Off
Glucose Alerts
•On
•Off
Off Set the limits using the
Glucose Limits feature.
Pump specifications194