3-2 LIFEPAK CR Plus Defibrillator Operating Instructions
Using the LIFEPAK CR Plus Defibrillator
WARNINGS AND CAUTIONS
To help ensure safe use of the LIFEPAK CR Plus Defibrillator, completely familiarize yourself with the
following warnings and cautions.
RESPONDING TO A CARDIAC ARREST PATIENT
If not treated, cardiac arrest will cause death. In a cardiac arrest situation, it is important to remember to
immediately call for help and activate your emergency response system.
Basic Steps for Using the LIFEPAK CR Plus Defibrillator
Responding to a cardiac arrest patient using the LIFEPAK CR Plus Defibrillator involves these basic
steps:
WARNINGS!
Shock hazard.
When instructed “Do not touch patient,” “Stand by,” or “Everyone clear,” remain still, do not touch the
defibrillator, patient, electrode pads or any material in contact with the patient. Make sure no one is
touching the patient when the defibrillator shocks the patient.
Shock hazard.
To remove an unwanted charge, disconnect the electrode cable from the defibrillator, wait for the
defibrillator to automatically remove the charge, or turn off the defibrillator.
Possible fire, burns, and ineffective energy delivery.
During defibrillation, material in contact with the electrode pads can cause electrical sparks, skin burns,
and divert important defibrillating energy away from the heart. Place electrode pads so that they adhere
to the skin completely. Do not allow the electrode pads to touch each other, medication patches,
dressings or any other material on the patient’s chest.
CAUTION!
Possible equipment damage.
Before using the defibrillator, disconnect all equipment from the patient that is not
defibrillator-protected.
Establish that the patient is in cardiac arrest, as instructed in
your CPR training. Place the LIFEPAK CR Plus Defibrillator near
the patient and press the
LID RELEASE/ON-OFF button to open the
lid and turn on the defibrillator. The voice prompt sequence
begins, guiding you through the response steps.